Gardner LCM IV User Manual

Page of 110
73
Standard programs
5.7.1.2 Take single wavelength readings (single reading)
1.
Insert the zero solution cell into the cell compartment.
2.
Press Zero.
Note: The Read key is only active after the zero reading has been 
completed.
3.
Insert the analysis cell into the cell compartment.
4.
Press Read
5.
For data storage, refer to 
5.7.2 Multi Wavelength mode – readings with more than one wavelength
In the Multi Wavelength mode, absorbance values can be 
measured at up to four wavelengths and the results can be 
mathematically processed to obtain sums, differences and 
relationships.
Absorbance readings: The light absorbed by the sample is 
measured in absorbance units.
% Transmittance measures the percent of the original light 
that passes through the sample and reaches the detector.
Turning on the concentration factor allows selection of a 
specific multiplier for converting absorbance readings to 
concentration. In a graph of concentration versus the 
absorbance, the concentration factor is the slope of the line. 
Concentration is calculated using a single factor for each 
wavelength, which is input by the user.
5.7.2.1 Set the Reading mode at different wavelengths 
In the main menu, press the Multi-Wavelength menu option. 
Press Options for Parameter Setup.
Table 8 Multi wavelength setup options
Options
Description
More
For further Options
Save icon
Symbol: Store Data, if Instrument Setup > Data Log Setup > Auto Store: Off is selected.
Symbol: Recall Data, if Instrument Setup > Data Log Setup > Auto Store: On is selected.
% Trans/Abs
To switch to % transmittance, concentration or absorbance readings