Gardner LCM IV User Manual

Page of 110
89
Advanced Operations
7.
Remove any present cuvettes from the cuvette 
compartment.
8.
Close the cuvette compartment cover.
9.
Press Start.
10.
Follow the instructions on the screen and insert the different 
filters one after another in the sequence shown. 
11.
Press Next after inserting a filter.
After the last reading the results are displayed.
12.
Press PC & Printer icon to send the data to a USB memory 
stick, PC or to a printer.
The files will be stored automatically as CSV files (Comma 
Separated Value). The file name will be formatted as 
"Verification.csv".
6.1.4
AQA Analytical Quality Assurance
6.1.4.1
Standard Configuration
Analyzing an AQA standard solution enables you to check your 
working procedure, photometer and accessories.
1.
Under "System Check", select AQA
or 
press AQA  in the toolbar on the right of the screen.
A list of standard solutions opens.