Gardner LCM IV User Manual

Page of 110
98
Maintenance
7.1.3
Cuvettes/cells
1.
After use, clean the glass cells with cleaning agents.
2.
Afterwards, rinse the cuvettes/sample cells several times 
with tap water and then thoroughly with deionized water.
Note: Glass cuvettes that contained organic solvents (such as 
chloroform, benzene, toluene etc.) must be rinsed out with acetone 
before being treated with cleaning agent. In addition, another rinse with 
acetone is necessary as a final treatment step before the 
cuvettes/sample cells are dried.
7.2
Lamp replacement
1.
Switch off instrument.
2.
Remove the power cable from the instrument.
3.
Turn the instrument to allow work access to the underside.
4.
Use a slotted screwdriver to loosen the screws on the cover 
of the lamp compartment.
5.
Remove the cover.
C A U T I O N
Potential Chemical/ Biological Exposure Hazards.
User correct laboratory practices if a risk of chemical exposure exists.
C A U T I O N
To avoid the risk of electric shock, disconnect the instrument from the 
power source before the lamp replacement procedure is commenced. 
C A U T I O N
Burn hazard. 
Wait until the lamp cools down. Contact with the hot lamp can cause 
burns.