Gardner LCS IV User Manual

Page of 110
7
Section 1
Specifications
These are subject to change without notice!
Performance specifications
LCS IV
LCM IV
Display mode
Color measurement, color difference 
measurement, absorbance and 
concentration
Color measurement
Color measurement
26 color ratios
5 color ratios
Colorimetric evaluation
All visual color ratios are calculated for standard light chart C and 2° standard 
observers in accordance with DIN 5033. Colorimetric color values can be switched 
to light type A, C, D65 and 2° or 10° standard observers.
Source lamp
Halogen lamp
Wavelength range
320–1100 nm
Wavelength Accuracy
± 1.5 nm (wavelength range 340–900 nm)
Wavelength reproducibility
≤ 0.1 nm
Wavelength resolution
1 nm
Wavelength calibration
Automatic
Wavelength range for color 
measurement
380–720 nm steps of 10 nm
Scanning speed
≥ 8 nm/sec (in steps of 1 nm)
Spectral bandwidth
5 nm 
Photometric measuring range
± 3 Abs (wavelength range 340–900 nm)
Photometric accuracy
5 mAbs at 0.0–0.5 Abs, 
1 % at 0.50–2.0 Ext 
Photometric linearity
< 0.5 % to 2 Abs
≤ 1 % at > 2 Abs with neutral glass at 546 nm 
Stray light
< 0.1 % T at 340 nm with NaNO
2
Data log
3000 color measurements, 
100 color reference values, 
1000 photometric measurements, 
20 wavelength scans, 20 time scans
400 color measurements
Physical and environmental specifications
Width
350 mm (13.78 in)
Height
151 mm (5.94 in)
Depth
255 mm (10.04 in)
Earth
4200 g (9.26 lb)
Ambient operating requirements
10–40 °C (50–104 °F), maximum 80 % relative humidity (without condensate 
formation)
Ambient storage requirements
–40–60 °C (–40–140 °F), maximum 80 % relative humidity (without condensate 
formation)
Additional technical data
Power connector via external power 
supply
Input: 100–240 V/47–63 Hz
Output: 15 V/40 VA