Equivalent Front-Panel Key Commands
INSTRUMENT STATE Keys
ADRS
The ADRS key displays the Test Set’s GPIB address in the upper left-hand corner
of the CRT. There is no equivalent GPIB command for the ADRS key. The current
address can also be viewed by looking at the
of the CRT. There is no equivalent GPIB command for the ADRS key. The current
address can also be viewed by looking at the
HP-IB Adrs
field on the I/O
CONFIGURE screen.
The Test Set’s GPIB address can be changed through GPIB by using the
:CONFigure:BADDress commands. If the Test Set’s GPIB address is changed
programmatically, all future GPIB commands must use the new address.
:CONFigure:BADDress commands. If the Test Set’s GPIB address is changed
programmatically, all future GPIB commands must use the new address.
Syntax
CONFigure:BADDress <integer number>
Example
OUTPUT 714;"CONF:BADD 15"
This sets the Test Set’s GPIB address to 15.
The Test Set’s GPIB address can be queried through GPIB by using the
:CONFigure:BADDress? commands.
:CONFigure:BADDress? commands.
Syntax
‘CONFigure:BADDress?
Example
OUTPUT 714;"CONF:BADD?"
ENTER 714;Address
This queries the Test Set’s GPIB address.