Chapter 6, Memory Cards/Mass Storage
Default Mass Storage Locations
Default Mass Storage Locations 
Built-in IBASIC Controller 
The default mass storage location for the built-in IBASIC Controller is the front 
panel memory card slot (mass storage volume specifier ":INTERNAL,4") after 
any of the following conditions: 
power-up 
initializing RAM with the SERVICE screen’s 
RAM Initialize
 function 
The mass storage location for the built-in IBASIC Controller can be changed 
using the MASS STORAGE IS command. Refer to the Instrument BASIC Users 
Handbook
 for further information on the MASS STORAGE IS command. 
Save/Recall Registers 
The default mass storage location for the Save/Recall registers is the Test Set’s 
internal RAM (no mass storage volume specifier) after any of the following 
conditions: 
power-up 
initializing RAM with the SERVICE screen’s 
RAM Initialize
 function 
resetting the Test Set using the front-panel PRESET key 
resetting the Test Set using the *RST GPIB Common Command 
The mass storage location for Save/Recall registers can be changed using the 
Save/Recall
 field in the I/O CONFIGURE screen. The default mass storage 
volume specifiers for the Save/Recall register mass storage locations are as 
follows: 
Internal selection - (no mass storage volume specifier, registers are saved to allocated 
RAM space) 
Card selection (not changeable) - ":INTERNAL,4" 
RAM selection (not changeable) - ":MEMORY,0,0" 
Disk selection - the 
External Disk Specification
 field in the TESTS 
(External Devices) screen.