Chapter 8, Programming the Call Processing Subsystem
Programming the CALL CONTROL Screen
Order
This field is used to send an order type mobile station control message on the 
forward voice channel to the mobile station. The orders available are: 
Change Power to Power Level 0 - 7
Maintenance (put the mobile station in maintenance mode) 
Alert (alert the mobile station)
The ORDER field is updated using the command:
:ORDER for system types AMPS, TACS, AND JTACS. This command is used to send 
an order type mobile station control message to the mobile station. The 
Access
 
annunciator will light momentarily while the Test Set is sending the mobile station 
control message.
A mobile station must be actively connected on a voice channel to the Test Set (that is, 
the 
Connect
 annunciator lit) before attempting to send an order to a mobile station.
The query form of the command (that is, :ORDer?) can be used to determine the last 
order sent to the mobile station using the :ORDer command.
Syntax
:ORDer <’><order message><’>
:ORDer?
Example
OUTPUT 714;"CALLP:ORD ’CHNG PL 0’"
OUTPUT 714;"CALLP:ORD?"
ENTER 714;Last_ord_sent$ 
OUTPUT 714;"CALLP:ORD?"