Chapter 8, Programming the Call Processing Subsystem
Programming the CALL BIT Screen
Reading the CALL BIT Screen Message Fields
This section provides programming information on how to read the contents of 
individual fields in the signaling messages available on the 
CALL BIT
 screen.
The syntactical structure for reading the contents of one or more fields from an 
individual message is as follows:
General Syntax
CALLP:<message name>:<field name><?>[<;><additional field><?>]
 lists the message names used to access each of the signaling 
messages available on the 
CALL BIT
 screen.
Example of Querying A Single Field
OUTPUT 714;"CALLP:MESS ’SPC WORD1’"
OUTPUT 714;"CALLP:SPOM1:SID?"
ENTER 714;Sid$ 
PRINT Sid$ 
Example Printout
"00000001110011"
Table 50
CALL BIT Screen Signaling Message Names
Message 
Message Name
SPC WORD1
SPOM1/SPOMESSAGE1
SPC WORD2
SPOM2/SPOMESSAGE2
ACCESS
ACCess
REG INC
RINCrement
REG ID
RIDentify
C-FILMESS
CFMessage
MS WORD1
MSWord
MSMessOrd
MSORder
MS IntvcH
MSVoice
FVC O Mes
FVORder
FVC C Mes
FVVoice