Agilent Technologies A.18.00 User Manual

Page of 581
62
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb
 
Chapter 2, Methods For Reading Measurement Results
HP
®
 BASIC ‘ON TIMEOUT’ Example Program
Comments for Recommended Routine
Table 7
Comments for Measure Function from ON TIMEOUT 
Example Program
Program Line 
Number
Comments
50
Send a Selected Device Clear (SDC) to the Test Set to put the GPIB subsystem 
into a known state. This allows the control program to regain programmatic 
control of the Test Set if it is in an error state when the program begins to run.
60
Command the Test Set to abort the currently executing measurement cycle. This 
will force the Test Set to stop waiting for any measurement results to be available 
from measurements which may be in an invalid state when the program begins to 
run.
290
Turn event initiated branches off (except ON END, ON ERROR and ON 
TIMEOUT) to ensure that the Measure function will not be exited until it is 
finished.
300
Set up a timeout for any I/O activity on the GPIB. This will allow the function to 
recover if the bus hangs for any reason.
310
Set the triggering mode to single followed by a trigger immediate command. This 
ensures that a new measurement cycle will be started when the TRIG:IMM 
command is sent. This sequence, that is: set to single trigger and then send a 
trigger command, guarantees that the measurement result returned to the ENTER 
statement will accurately reflect the state of the DUT when the TRIG:IMM 
command was sent. The ’IMM’ keyword is optional.
320
Send the query command passed to the Measure function to the Test Set.
330
Read the measurement result.
340
Set the trigger mode to repetitive retriggering. Setting the trigger mode to 
repetitive will be implementation dependent.
350
Re-enable event initiated branching. If any event initiated branches were logged 
while the Measure function was executing they will be executed when system 
priority permits.