Emerson 396P User Manual

Page of 54
Maintenance
37
Section 5: Maintenance
5.1 
General Information
The 396P and 396PVP Sensors require minimum maintenance. The sensor should be kept clean
and free of debris and sediment at all times. The frequency of cleaning by wiping or brushing with
a soft cloth or brush is determined by the nature of the solution being measured. The sensor
should be removed from the process periodically and checked in buffer solutions.
If the sensor will not calibrate, refer to your analyzer/ transmitter instruction manual for proper
test procedures. If it is determined that the sensor has failed, it 
5.2 
Automatic Temperature Compensator
The temperature compensator element is temp erature sensitive and can be checked with an
ohmmeter. Resistance increases with temperature.
The 3K element will read 3000 ohms ±1% at 25°C (77°F) and a Pt100 will read 110 ohms. Resist-
ance varies with temperature for a 3K and Pt100 element and can be determined according to
Table 6-2 or the following formula: 
R
T
=R
O
[l+R
1
(T-20)] 
Where R
T
= Resistance
T = Temperature in °C
Refer to Table 6-1 for R
O
and R
1
values
5.3 
396P & 396PVP pH Sensors
5.3.1
Electrode Cleaning
If the electrode is coated or dirty, clean as follows:
1. Remove the sensor from process.
2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt
or coating, go to Step 3. (Detergents clean oil and grease; acids remove scale.)
3. Wash the glass bulb in a mild detergent solution and rinse it in clean water. If this does not
clean the glass bulb, go to Step 4.
4. Wash the glass bulb in a dilute 5% hydro chloric acid solution and rinse with clean water.
Soaking the sensor overnight in the acid solution can improve cleaning action.
DANGER
BEFORE REMOVING THE SENSOR, be absolutely certain that the process pressure is reduced to 0 psig
and the process temperature is lowered to a safe level!
CAUTION
The solution used during the following check is an acid and should be handled with care. Follow the
directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution
come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
396P + 396PVP Sensors Instruction Manual
Section 5: Maintenance
LIQ_MAN_396P_396PVP
August 2013