Fisher 244ER User Manual

Page of 64
2-11
Screen Conventions
The Models 244EH and 244ER Configuration Software follows the 
Microsoft Windows screen conventions with the following exception: 
Configuration parameters change color to indicate fields where 
information has changed. Before placing a new transmitter into service, 
or before returning a transmitter to service after changing 
configuration information, verify that the values in the fields that have 
been changed reflect the correct configuration parameters for your 
application. For example, if you change the Sensor Type field to 
PT100–
Alpha 392
, the Number of WiresUnits4 mA Point, and 20 mA Point 
fields all change to reflect the default 
PT100–Alpha 392
 sensor values. 
Verify all information before placing the transmitter into service.
Model 244EC Configuration
Interface
The Model 244EC Configuration Interface is a portable, self-contained 
link between your PC and a Model 244. The Model 244EC connects to a 
PC serial port with a standard 9-pin interconnecting plug and connects 
to a transmitter with two MINIGRABBER
 clips. 
The Model 244EC will also operate using a wall power adapter or a 
single replaceable 9-volt battery. 
Setting Up the Model 244EC Configuration Interface
All necessary power is provided through the configuration leads from 
the Model 244EC to the transmitter. The sensor does not need to be 
disconnected in order to configure the transmitter.
To set up the Model 244EC Configuration Interface and prepare it for 
use, refer to Figure 2-10 on page 2-12 and follow the procedure 
described below.
1. Install the 9-volt battery in the Model 244EC. Be sure the power 
switch remains in the “OFF” position.
2. Attach the ribbon cable from the Model 244EC to the serial port of 
your PC using the 9-pin interconnecting plug. If your PC has a 
25-pin serial port, you will need a 25-pin to 9-pin adapter to 
accommodate the connection.
NOTE
It is not necessary to power down your PC before you attach the 
Model 244EC. However, if you are using a desktop model, you may wish 
to power down to reduce the risk of electric shock or computer damage.
3. Attach the configuration leads to the Model 244EC using the 
banana jacks provided. Be sure to observe proper polarity—attach 
the red lead to the positive (+) jack on the Model 244EC and the 
black lead to the negative (–) jack.
4. Attach the configuration leads to the configuration terminals 
(labeled “PROG”) on the transmitter using the MINIGRABBER 
clips provided. Be sure to observe proper polarity—attach the red 
lead to the positive (+) terminal on the transmitter and the black 
lead to the negative (–) terminal.