Fisher 244ER User Manual

Page of 64
Rosemount Models 244EH and 244ER PC-Programmable Temperature Transmitters
2-16
Figure 2-13. Transmitter Power,
Sensor, and Configuration Terminals
FAILURE MODE
The Models 244EH and 244ER features software driven alarm 
diagnostics and an independent circuit. These features are designed to 
provide separate backup alarm output in case the microprocessor, 
electronics, hardware, or software fails. The alarm levels are user 
selectable using the Failure Mode Switch. The position of the jumper 
determines the direction in which the output is driven (HI or LO) in 
case of alarm. The jumper switch feeds into the Digital-to-Analog (D/A) 
converter, which drives the proper alarm output even if the 
microprocessor fails.
The values to which the transmitter drives its output in failure mode 
depend on whether it is factory configured to standard or 
NAMUR-compliant operation. See “Failure Mode” on page A-2 for 
these parameters.
To determine the failure mode configuration of a transmitter, review 
the low and high alarm levels on the 
SERVICE
 menu provided by the 
configuration software.
Changing Switch Positions
To change the failure mode on the Model 244EH or 244ER transmitter, 
follow the steps described below.
1. If the transmitter is mounted in an enclosure, remove the 
enclosure cover. 
2. Locate the orange failure mode switch. The switch is located near 
the power terminals on the Model 244EH and in the center of the 
front panel on the Model 244ER (see Figure 2-13).
3. Move the switch to the desired alarm setting. To set the failure 
mode to high alarm, position the switch toward the “HI” mark on 
the terminal block; to set the failure mode to low alarm, position 
the switch in the opposite direction.
4. Replace the enclosure cover (if applicable). Enclosure covers must 
be fully engaged to meet explosion-proof requirements.
Model 244EH
Model 244ER
1
2
3
4
Failure Mode
Switch
Power Terminal
Sensor
Terminals
Configuration
Terminals
2
44-1
260B
02D
,
110
5E
01B
Failure Mode Switch
Sensor Terminal
Power/Configuration
Terminals