Setting Up the Infiniium DCA
Cleaning Connectors
duce a good return-loss measurement. The quality of the polish establishes 
the difference between the “PC” (physical contact) and the “Super PC” con-
nectors. Most connectors today are physical contact which make glass-to-glass 
connections, therefore it is critical that the area around the glass core be clean 
and free of scratches. Although the major area of a connector, excluding the 
glass, may show scratches and wear, if the glass has maintained its polished 
smoothness, the connector can still provide a good low level return loss con-
nection.
If you test your cables and accessories for insertion loss and return loss upon 
receipt, and retain the measured data for comparison, you will be able to tell in 
the future if any degradation has occurred. Typical values are less than 0.5 dB 
of loss, and sometimes as little as 0.1 dB of loss with high performance con-
nectors. Return loss is a measure of reflection: the less reflection the better 
(the larger the return loss, the smaller the reflection). The best physically 
contacting connectors have return losses better than 50 dB, although 30 to 
40 dB is more common.
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections 
on the cable end face can be detected as well as cracks or chips in the fiber 
itself. Use a microscope (100X to 200X magnification) to inspect the entire 
end face for contamination, raised metal, or dents in the metal as well as any 
other imperfections. Inspect the fiber for cracks and chips. Visible imperfec-
tions not touching the fiber core may not affect performance (unless the 
imperfections keep the fibers from contacting).
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber-
optic cables and Agilent Technologies universal adapters. The initial cleaning, 
using the alcohol as a solvent, gently removes any grit and oil. If a caked-on 
layer of material is still present, (this can happen if the beryllium-copper sides 
of the ferrule retainer get scraped and deposited on the end of the fiber during 
insertion of the cable), a second cleaning should be performed. It is not 
uncommon for a cable or connector to require more than one cleaning.
C A U T I O N
Agilent Technologies strongly recommends that index matching compounds 
not be applied to their instruments and accessories. Some compounds, such as 
gels, may be difficult to remove and can contain damaging particulates. If you 
think the use of such compounds is necessary, refer to the compound 
manufacturer for information on application and cleaning procedures.