Agilent Technologies 86100b User Manual

Page of 90
3-7
Using the Infiniium DCA
Eye/Mask Mode
Eye/Mask Mode
The Eye/Mask mode allows you to per-
form NRZ (Non-Return to Zero), or RZ 
(Return to Zero) eye diagram measure-
ments and eye mask tests. The eye dia-
gram is typically produced by triggering 
the instrument with a synchronous 
clock signal. 
Eye/Mask mode measurements only 
work when an eye diagram, and not a 
pulse, is present on the screen. Mea-
surements made on a pulse waveform 
while in Eye/Mask mode will fail.
Once you are in Eye/Mask mode, press the Autoscale button. This ensures 
that an optimum eye diagram is displayed on the graticule. An optimum eye 
diagram consists of a full display of the eye in addition to portions of the wave-
form preceding and following the eye. 
When you select the Eye/Mask mode, the 
measurement toolbars shown here appear. 
These measurement toolbars include mea-
surements such as rise time, fall time, jitter, 
and extinction ratio, as well as the mask test 
start and stop controls. 
Eye measurement and mask test results 
appear below the display graticule. A maxi-
mum of four eye measurement test results 
are displayed and include any failures or 
instrument messages that may occur.
To learn more about the Eye/Mask mode and 
switching from NRZ to RZ measurement 
mode, consult the Help.