Symmetricom 5402 User Manual

Page of 40
Operating Manual
Operation
PN 8500-0150, Rev. A, May 2005
18
Response Format:
Description: 
Nn - Start-up Index; n (1 to 6) identifies which one of 6 elements in the start-up table was 
reported. 
Ggg.g - Dynamic Gain Value; gg.g (0 to 99.9) represents the dynamic gain value that will be 
used during start-up for element "n". 
Ttttttt - End Time; tttttt is the end time of element "n". The next element in the start-up table 
will be used after this time. The time is the number of times the discipline loop has been run. 
Using a time of "000000" will cause that element to be skipped. 
Calibrate Unit 
Command Format: 
Description: 
Commands a self-calibration. Before commanding a Self-Cal, connect one of the 1 PPS 
outputs to the 1 PPS input using a short coax cable. The unit will be set to provide an output 
1 PPS based on the electrical length of the cable used to connect the output to the input. The 
maximum cable delay is limited to 1 microsecond. 
Response Format:
Note: If element 1 is used (not 000000), then a 1 PPS resync will occur before element 2 
starts. After all 6 elements have been completed, and before normal operation begins, 
the discipline filter will be reinitialized. 
Note: The "Calibrate Status" command should be used to verify that the calibration has 
completed (takes several minutes). The programmable delay value, that controls the 
phase relationship of the 1 PPS to the 10 MHz, should be set prior to running "Calibrate 
Unit".