Syntax and Commands
PS2520G and PS2521G Programmer Manual
31
STATus:OPERation:INSTrument:ENABle <NRf>
STATus:OPERation:INSTrument:ENABle?
<NR1> is an integer from 0 to 32767.
<NR1>
STATus:OPERation:INSTrument:ENABle 32767 sets all 15 bits of
the register high.
STATus:OPERation:INSTrument:ENABle? returns 0 if all 15 bits of
the register are low.
STATus:OPERation:INSTrument[:EVENt]?
Returns and clears the contents of the OPERation INSTrument
register. The response is a decimal value that summarizes the binary
values of the set bits.
STATus:OPERation:INSTrument[:EVENt]?
<NR1>
STATus:OPERation:INSTrument:EVENt? returns 0.
STATus:OPERation:INSTrument:ISUMmary<n>:CONDition?
(Query Only)
Returns the contents of the OPERation INSTrument ISUMmary<n>
register. The PS2520G and PS2521G Programmable Power Supplies,
however, do not use the OPERation INSTrument ISUMmary<n>
register to report any conditions.
Syntax
Parameters
Returns
Examples
Syntax
Returns
Examples