Tektronix TDS460A User Manual

Page of 433
Command Descriptions
2–110
TDS Family Oscilloscope Programmer Manual
:
DIAg
:
?
LOG
RESUlt
<QString> in the following format:
<Status>,<Modulename>[,<Status>,<Modulename>...]
DIAG:RESULT:LOG?
might return 
:DIAG:RESULT:LOG "pass--Processor,pass--Display,
pass--FP/Proc Interface,FAIL--Front Panel"
DIAg:SELect:ACQUISition (No Query Form)
Selects the acquisition system test sequence that will be run when the
DIAg:STATE EXECUte command is sent. This command is equivalent to setting
Area in the Utility menu when System is set to Diag/Err.
Calibration and Diagnostic
DIAg:SELect:ACQUISition ALL
:
SELect
DIAg
:
ACQUISition
<Space>
ALL
ALL selects functional, memory, and register tests.
DIAg:SELect:ALL (No Query Form)
Specifies that all system test sequences will be run when the DIAg:STATE
EXECUte command is sent. This command is equivalent to setting Area in the
Utility menu when System is set to Diag/Err.
Calibration and Diagnostic
DIAg:SELect:ALL ALL
:
SELect
DIAg
:
<Space>
ALL
ALL
ALL selects functional, memory, and register tests for the acquisition, processor
and display systems, and self diagnostics for the front panel.
Returns
Examples
Group
Syntax
Arguments
Group
Syntax
Arguments