Tektronix TDS460A User Manual
Command Descriptions
2–56
TDS Family Oscilloscope Programmer Manual
NOTE. If you set the acquisition system to single sequence, envelope mode, and
set the number of envelopes to infinity, the digitizing oscilloscope will envelope a
maximum of 2001 acquisitions.
set the number of envelopes to infinity, the digitizing oscilloscope will envelope a
maximum of 2001 acquisitions.
ACQUIRE:STOPAFTER RUNSTop
sets the oscilloscope to stop acquisition when the user presses the front-panel
RUN/STOP button.
sets the oscilloscope to stop acquisition when the user presses the front-panel
RUN/STOP button.
ACQUIRE:STOPAFTER?
might return
might return
SEQUENCE.
ALIas
Turns command aliases on or off. This command is identical to the
ALIas:STATE command.
ALIas:STATE command.
Alias
ALIas { OFF | ON | <NR1> }
ALIas?
ON
ALIas
<Space>
?
<NR1>
OFF
OFF or <NR1> = 0 turns alias expansion off. If a defined alias label is sent when
ALIas is OFF, an execution error (110, “Command header error”) will be
generated.
ALIas is OFF, an execution error (110, “Command header error”) will be
generated.
ON or <NR1> 0 turns alias expansion on. When a defined alias is received, the
specified command sequence is substituted for the alias and executed.
specified command sequence is substituted for the alias and executed.
ALIAS ON
turns the alias feature on.
turns the alias feature on.
ALIAS?
returns
returns
1 when aliases are on.
Examples
Group
Syntax
Arguments
Examples