Table 2–28: Offset Ranges for the TDS 400A, 510A, 540C, 600B, 754C, & 784C (All
Channels) and the TDS 520C & 724C (Channel 1 & Channel 2) using a 1x
Probe 
CH<x>:SCAle
OFFSet Range
1 mV/div – 99.5 mV/div
±
1 V
100 mV/div – 995 mV/div
±
10 V
1 V/div – 10 V/div
±
100 V
Table 2–29: Offset Ranges for the TDS 520C & 724C (Aux 1 & Aux 2) using a 1x
Probe 
CH<x>:SCAle
OFFSet Range
50 mV/div & 100 mV/div
±
.5 V
500 mV/div & 1 V/div
±
5.0 V
5 V/div & 10 V/div
±
50 V
CH1:OFFSET 0.5E+00
lowers the channel 1 displayed waveform by 0.5 volts.
CH1:OFFSET?
might return 
500.0E-3, indicating that the current channel 1 offset is 0.5 volts.
CH<x>:POSition
Sets or queries the vertical position of the specified channel. The position value
is applied to the signal before digitization. This is equivalent to setting Position
in the Vertical menu or adjusting the front-panel Vertical Position knob.
Vertical
CH<x>:OFFSet
CH<x>:POSition <NR3>
CH<x>:POSition?
Examples
Group
Related Commands
Syntax