Tektronix TDS420A User Manual
Command Descriptions
2–234
TDS Family Oscilloscope Programmer Manual
MEASUrement:REFLevel:ABSolute:MID2
Sets or queries the mid reference level for the “to” waveform when taking a
delay measurement, and is the 50% reference level when MEASUrement:RE-
FLevel:METHod is set to ABSolute. This command is equivalent to setting the
Reference Levels in the Measure menu.
delay measurement, and is the 50% reference level when MEASUrement:RE-
FLevel:METHod is set to ABSolute. This command is equivalent to setting the
Reference Levels in the Measure menu.
Measurement
MEASUrement:REFLevel:ABSolute:MID2 <NR3>
MEASUrement:REFLevel:ABSolute:MID2?
MID2
?
<Space>
<NR3>
MEASUrement
:
REFLevel
:
ABSolute
:
<NR3> is the mid reference level, in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:MID2 0.5
sets the mid reference level for the delay waveform to 0.5 volts.
sets the mid reference level for the delay waveform to 0.5 volts.
MEASUrement:REFLevel:METHod
Specifies which reference levels are used for measurement calculations. This
command is equivalent to setting the levels in the Reference Levels side menu.
command is equivalent to setting the levels in the Reference Levels side menu.
Measurement
MEASUrement:REFLevel:METHod { ABSolute | PERCent }
MEASUrement:REFLevel:METHod?
METHod
MEASUrement
:
REFLevel
:
<Space>
?
ABSolute
PERCent
Group
Syntax
Arguments
Examples
Group
Syntax