Tektronix TDS420A User Manual
Command Descriptions
2–314
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:SLEWRate:SLEWRate? (Query Only)
TDS 500C, 600B, & 700C Only
Returns the slew rate value. This is the
((Upper Threshold – Lower Threshold) / Delta Time)
The value is limited to the three most significant digits.
Trigger
TRIGger:MAIn:PULse:SLEWRate:SLEwrate?
TRIGger
:
:
PULse
MAIn
:
?
SLEWRate
:
SLEWRate
TRIGGER:MAIN:PULSE:SLEWRATE:SLEWRATE?
returns the slew rate. It is given as an appropriate amount of volts per second.
For example,
returns the slew rate. It is given as an appropriate amount of volts per second.
For example,
1.2E+6 would represent a slew rate of 1.2 megavolts/second.
TRIGger:MAIn:PULse:SLEWRate:THReshold:BOTh
TDS 500, 600B, & 700C Only
Sets the upper and lower slew rate trigger thresholds. This is equivalent to
selecting Thresholds from the main Trigger’s Slew Rate menu and entering a
value in the resulting side menu’s High Threshold or Low Threshold items.
selecting Thresholds from the main Trigger’s Slew Rate menu and entering a
value in the resulting side menu’s High Threshold or Low Threshold items.
Trigger
TRIGger:MAIn:PULse:SLEWRate:THReshold:BOTh { TTL | ECL }
THReshold
TRIGger
:
:
PULse
MAIn
:
:
BOTh
SLEWRate
ECL
TTL
TTL sets the upper threshold to 1.8 V and the lower threshold to 0.8 V.
ECL sets the upper threshold to –1.1 V and the lower threshold to –1.5 V.
Group
Syntax
Examples
Group
Syntax
Arguments