Tektronix TDS420A User Manual
Command Descriptions
2–318
TDS Family Oscilloscope Programmer Manual
<Space>
EITher
?
STAYSHigh
STAYSLow
TRIGger:
MAIn
POLarity
:
PULse
:
:
TIMEOut
STAYSLow (or NEGAtive) indicates that a pulse edge must stay low the required
time period for timeout triggering to occur.
time period for timeout triggering to occur.
STAYSHigh (or POSITIVe) indicates that a pulse edge must stay high the required
time period for timeout triggering to occur.
time period for timeout triggering to occur.
EITher indicates either STAYSHigh or STAYSLow polarity.
TRIGGER:MAIN:PULSE:TIMEOUT:POLARITY EITHER
specifies that the polarity of the timeout trigger can be either positive or
negative.
specifies that the polarity of the timeout trigger can be either positive or
negative.
TRIGger:MAIn:PULse:TIMEOut:TIMe
TDS 500C, 600B, & 700C Only
Sets or queries the pulse timeout trigger time, in seconds. This command is
equivalent to selecting Time in the Trigger menu and entering a value with the
keypad or general purpose knob.
equivalent to selecting Time in the Trigger menu and entering a value with the
keypad or general purpose knob.
Trigger
TRIGger:MAIn:PULse:TIMEOut:TIMe <NR3>
TRIGger:MAIn:PULse:TIMEOut:TIMe?
?
<Space>
<NR3>
TRIGger
:
:
PULse
MAIn
TIMEOut
:
:
TIMe
<NR3> is the timeout time period, in seconds.
Arguments
Examples
Group
Syntax
Arguments