Omega Multifunction Indicator-Simulator User Manual
24
9
OPERATION & APPLICATIONS
9.1
Power ON
A
TTENTION
:
ALL VALUES IN THE FOLLOWING FIGURES ARE ONLY LISTED AS AN EXAMPLE
.
During set-up and load memory remember that the instructions of the manual related to key operation have the following
meaning:
<A> + <B>
meaning:
<A> + <B>
Press the <A> key and keeping the pressure on the key, press then the <B> key.
<A> , <B>
Press in sequence, first the <A> key and then the <B> key.
•
To power the instrument press the <ON> key; the indication :
... CL526...
will appear for a few seconds.
•
The instrument runs an autodiagnostic routine for the self-checking of critical circuits and components.
•
A positive check will be shown with the indication
Test OK Ver 7.001
for about one second.
•
The number on the right side of the display indicates the version of the memory installed on the instrument.
•
The instrument is ready for measurement (IN mode) with the previously selected operating mode, as indicated below:
In 1280.6°C TcK:
Any faulty conditions that may be indicated are described in par. 8.9.
9.2
Battery voltage indication
•
To recall the battery voltage on the display press the <SHIFT> + <BATTERY> keys. The indication will be as
follows:
BAT 5.2 V IIII
•
The horizontal bar indicates the level of charge of the battery (each bar is equivalent to 25% of the full capacity)
•
The “low” limit of the battery voltage, for the correct operation of the instrument, is +4.6V.
•
Press any key to reset the operative mode.
•
During normal operating modes (measure or simulation), “low battery” condition will be shown as follows:
Out 1248.3°C TcK:
The battery symbol indicates that the battery has enough energy for about 30 minutes operation.
A black symbol means that battery charge is below minimum: batteries must be recharged.
A black symbol means that battery charge is below minimum: batteries must be recharged.
9.3
Operating mode set up
To select the required operating mode follow the procedures indicated below.