Omega Multifunction Indicator-Simulator User Manual

Page of 50
30
End 
Start 
   Time    Soak
    Time
    Soak
 Z step 
  To memory load the cycle parameters, follow the procedure indicated below
  Select the required tecnical unit or electrical parameters
  Select the required decimal point position
 Press 
<SHIFT> + <TIME>
 keys to enter the cycle set-up procedure obtaining the following indication
Time   0h  0m 50s  
related to the "time" in hours, minutes and seconds. The maximum setting is limited to 5 hours 33 minutes  20  seconds
(20.000 seconds)
 Press 
the <START> key to obtain the following indication:
Start       0.0mV  
  Press the <▲> and <▼> cursors to set the "Start" level of the cycle
 Press 
<ENTER>+<START> keys to memory store the new value
 Press 
the 
<END>
 key to obtain the following indication:
End       100.0mV  
 Press 
<▲> and <▼> cursors to set the "end" level of the cycle
 Press 
<ENTER>+<END> keys to memory store the new value
 Press 
the 
<STEP>
 key to obtain the following indication:
Step           1.0mV 
 Press 
<▲> and <▼> cursors to set the size of each individual step. To have a continuous ramp set the minimum
possible value (e.g. = 0.1)
 Press 
<ENTER>+<STEP> keys to memory store the new value
 Press 
the 
<TIME>
 key to obtain the following indication:
Time   0h  0m 50s  
 Press 
<▲> and
 <▼> cursors to set the required ramp time - max. 5h-33m-20s (20000 seconds)
 Press 
<ENTER>+<TIME> keys to memory store the new value
  A setting of 0h-0m-00s allows a manual step advance each time the <AUTORAMP> key is pressed.
 Press 
the 
<SOAK>
 key to read the following indication:
Soak   0h  0m 50s  
 Press 
<▲> and <▼> cursors to set the waiting time (or soak time). If the setting is 0h-0m-0s the waiting or soak time
is excluded