2.  Mounting the Flow Controller. 
 
The FMA 3200/3400 Series controllers have no particular orientation or 
installation requirements so may be mounted in any convenient position. 
 
It is recommended that units be fixed to a suitable substrate using the two 
4-40 mounting holes provided. 
 
 
Mounting View from Bottom 
(mounting hardware not included with sensor) 
 
 
 
3.  Tubing Connections 
 
All tubing must be clean, dry and purged with clean dry air before 
installation of the FLO-CONTROLLER
®
.  
 
If the gas to be used may contain particles then a filter (20 microns or 
less) should be installed upstream of (before) the unit.  
 
When connecting the sensor to the tubing, take care not to over-tighten 
the fittings or leaking may occur.  
 
 
Caution: Only use the fittings factory installed on the unit. If 
the fittings are removed the calibration of the unit may be 
effected and leaking may occur. If different fittings are required 
please contact the Omega Engineering Customer Service 
Department for assistance.