Omega CL526 User Manual
28
To select the required group follow the procedure indicated below:
•
Press
the
<SELECT>
key to obtain one of the menu pages
•
Press
the
<0>
key to obtain the following indication:
Sel STO Group: A
•
Press
the
<▲> or <▼> key to select the required group number
•
Press
the
<0>
key to confirm the selection and to return to the previously selected operative mode.
9.3.9.2
Data memory manual recall
To recall data memory values, select first the appropriate or required group number and press then the <0>, <1>, or <2>
key.
key.
9.3.9.3
Data memory automatic scanning
The 60 stored items of data can be assigned, besides the group number, to a pre-programmed sequence.
The programmed sequence (a number of calibration points, or a number of data items to be supervised), includes a
linear sequence of memory starting from a "start" point and ending with a "end" point".
Memory for seven different pre-programmed sequences are available and the procedure is indicated below.
The numerical value and the parameter indicated below are an operative example eg. to memory load 5 calibration
points of a scale of a potentiometric recorder (temperature with a thermocouple type K).
The programmed sequence (a number of calibration points, or a number of data items to be supervised), includes a
linear sequence of memory starting from a "start" point and ending with a "end" point".
Memory for seven different pre-programmed sequences are available and the procedure is indicated below.
The numerical value and the parameter indicated below are an operative example eg. to memory load 5 calibration
points of a scale of a potentiometric recorder (temperature with a thermocouple type K).
1 = -50°C
2 = 0°C
3 = +100°C
4 = +200°C
5 = +500°C
2 = 0°C
3 = +100°C
4 = +200°C
5 = +500°C
As described in par. 8.6.9.1 memory store the calibration data as follows:
point 1 =
memory 0 - group A
point 2 =
memory 1 - group A
point 3 =
memory 2 - group A
point 4 =
memory 0 - group B
point 5 =
memory 1 - group B
The simulation program set-up procedure is now required.
•
Press
the
<SELECT>
key to obtain one of the menu pages
•
Press
the
<1>
key to obtain the following indication:
SEL Program # 1
•
Press
the
<▲
▲
▲
▲>
and <▼
▼
▼
▼>
keys to enter the program number required (1 to 7)
•
Press
the
<1>
key to confirm the selection of the program number and to return to the indication of one of the menu
pages
•
Press
the
<2>
key to obtain the following indication:
Prog #1 from A1
•
Press the <▲>
and <▼> keys to select the "start" group/memory (eg. A1)
•
Press
the
<2>
key to confirm the "start" point obtaining the following indication:
Prog #1 to A2
•
Press
<s>