Omega CL526 User Manual

Page of 50
32
RJ :     0.0°C ext 
is displayed, it means that an external reference junction compensation has been selected for a temperature of 0.0°C
(programmable from -50°C to +100°C).
The reference junction compensation mode can be reprogrammed as indicated in par. 8.5.6. The external reference
junction compensation value can be programmed as indicated in par. 8.5.13.2.
9.3.12
Scale factor program
The “scale factor” mode is a method to read or to simulate electrical signals values in terms of engineering units.
The example explains the procedure of installing the “scale factor” function for the calibration of a potentiometric recorder
with a scale from 0.0 mbar to 400.0 mbar corresponding to the required electrical linear input signal.
 Press 
<SHIFT> + <PROGRAM X> keys to enter the "Scale Factor" set-up procedure. The display will indicate the low
end of the scale eg. in mbar.
LO:       0.0 Prog 
  Press one of  the
 <!
!
!
!> or <"
"
"
"> keys if a  decimal point shift is required.
 Press 
the
  <
> or  <
> keys to adjust to the required value.
 Press 
the 
<ENTER>
 key to load in the memory the value and to advance the program one step: the display will
indicate the full scale value of the technical unit (eg. mbar).
HI:     400.0 Prog 
 Press 
the 
<
>
 or <
>
 keys to adjust the full scale value.
 Press 
the 
<ENTER>
 key  to load in the memory the value: the display will indicate one of the menu pages as follows:
Type : 0-1000  mV  
Type :  0-100  mV  
Type :   0-10   V  
Type :    1-5   V  
Type :  0-400      
Type :   4-20  mA  
Type :   0-20  mA  
 Select, 
through 
<
>
 or  <
>
 keys, the required page and
 press 
the 
<ENTER>
 key to  memory load the needed parameter. The display will indicate one of the two following
pages:
Mode :    Linear 
Mode :    Square