Omega 3400 User Manual

Page of 26
M-4271/0707, pg. 19 of 26 
E.
 
Specifications 
 
Series 
FMA 3200 
FMA 3400 
FMA 3200ST 
FMA 3400ST 
Accuracy  
(including linearity) 
±1.5% of 
Full Scale* 
±1.5% of 
Full Scale* 
Second and 
third gases 
±3.0% F.S.* 
±1.5% of 
Full Scale* 
±1.5% of 
Full Scale* 
Second and 
third gases 
±3.0% F.S.* 
Repeatability ±0.25% 
Full 
Scale* 
Pressure Rating 
150 psig (10.3 bar) 
500 psig (34.5 bar) 
Pressure Sensitivity 
±0.02% Full Scale* per psi (per 69 mbar) 
Temperature Rating 
Operating Range: 5 to 55ºC  
Recommended Range (for best performance) : 10 to 40ºC 
Storage Range: 0 to 70ºC 
Temperature 
Sensitivity 
±0.15% F.S.* or less per ºC 
Valve Normally 
Closed 
Positive Shut-off up to 60psig (4 bar) 
Body Leak Integrity 
1x10
-7
 sccs of He 
Wetted Materials 
Aluminum 
304 Stainless Steel 
316 Stainless Steel 
303 Stainless Steel 
304 Stainless Steel 
316 Stainless Steel 
Epoxy 
O-Ring Material 
Viton
®
 
Fitting Material 
Choose from acetal, brass, or stainless steel 
Recommended 
Filtration 
20 microns or less 
Optional inline filters available 
Compatible gases 
Clean, dry gases compatible with wetted materials 
Output Signal 
0-5VDC, Impedance greater than 2.5 KΩ  
External Set-point 
Signal 
0-5VDC, Integrated 2MΩ load  
Internal Set-point 
Signal  
N/A Front 
Panel 
Adjustment 
N/A Front 
Panel 
Adjustment  
Warm-Up Time 
Less than 5 minutes 
Integrated Display 
N/A 
3½ digit 
N/A 
3½ digit 
Typical Power 
Consumption 
Standard: 12 VDC @ 250 mA (12.5-15 VDC) 
“E” Suffix: 24 VDC @ 130 mA (22-25 VDC) 
Peak Power 
Consumption 
Standard: 12 VDC @ 400 mA (12.5-15 VDC) 
“E” Suffix: 24 VDC @ 260 mA (22-25 VDC) 
Electrical 
Connections 
Integrated 36” (92 mm) cable, terminated with: 
Standard: 6-pin Mini-DIN male (PS/2 Style) 
D1 Option: 9-pin D-Sub male 
D2 Option: 15-pin D-Sub male 
Certifications CE 
Approved 
89/336/EEC (EN 55011 & EN 50082-1) 
73/23/EEC Low Voltage Directive 
*Specifications from 10-100% of rated flow. Linearity is best fit straight line. All calibrations 