Omega 3400 User Manual

Page of 26
M-4271/0707, pg. 23 of 26 
H.
 
Troubleshooting Guide 
 
Symptom 
Possible Cause 
Method of Correction 
Clicking noise from 
controller 
Unit in error mode 
Check there is sufficient 
pressure and that the flow 
path is not restricted or 
blocked 
Unit wired incorrectly 
Check wiring is according to 
Section B5 
Loose connection 
Check all connectors and 
wiring 
Damaged connector pins 
Contact Omega Engineering 
Blocked flow path 
Check flow path for 
obstructions.  
Piping leak before sensor 
Check all piping and 
connections. 
Insufficient power 
Check the power supply 
output and increase if 
necessary 
Output load resistance too low 
Ensure the voltmeter or 
data acquisition system or 
display has an impedance of 
greater than of 2.5kohm 
Flow too low for the unit 
Ensure that the flow being 
measured is within the 
capabilities of the unit 
No response 
Unit damaged or faulty 
Contact Omega Engineering 
Particles in flow path 
Add filtration before the 
sensor. 
Flow path obscured 
Remove any debris or 
blockage in the flow path 
eg. PTFE tape. 
Unit calibrated for a different 
gas 
Check calibration certificate 
and apply a “K” Factor to 
readings if necessary. 
Gas composition is variable 
Contact Omega Engineering 
Fittings have been changed 
Replace the factory installed 
fittings 
Moisture in gas 
Ensure gas is clean and dry 
Insufficient warm-up period 
Allow the unit to warm-up 
for at least 5 minutes. 
Zero drift 
Verify the zero and adjust 
as necessary as explained in 
Section C  
The gain potentiometer has 
been adjusted 
Contact Omega Engineering  
Unit needs recalibration 
Contact Omega Engineering 
Inaccurate control 
Flow too high for the unit 
Ensure that the flow being 
measured is within the 
capabilities of the unit