I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
from the cell before entering this command since it will perform the shunt calibration 
measurement that is recorded for use with the calibration check command.  The format 
is CTP0. 
 
Example: 
@123CTP0 
Acknowledge: 
@123 Calibrate Command - Reading for Shunt Check... 
          [10 seconds delay] 
@123 Calibrate Command Completed
 
     Ch A = S/N 123456, 100.00 LbI       , 3.00150 mV/v,  
                                          3.00230 mV/v, 
                                          3.00238 mV/v,  
                                          3.00353 mV/v, 
            10.00 V   ,  Cal on Oct27-99,  49.532 LbI  Shunt 
     unused S/N  13368,  10.00 Lb       , 3.00121 mV/v,  
            10.00 V,     Cal on Oct27-99, 4.9292 Lb Shunt 
     unused S/N  63220, 100.00 Lb      , 2.99984 mV/v,  
            10.00 V   , Cal on Oct27-99, 49.381 Lb  Shunt 
     Ch B = S/N  89991, 1000.0 Lb      , 4.50015 mV/v,  
             5.00 V   , Cal on Oct27-99, 486.45 Lb  Shunt 
 
Calibrate by Shunt Command (CS) 
The Calibrate by Shunt (CS) command is used to calibrate the cell using the shunt 
calibration factor expressed in pounds (or pound-inches for a torque cell).  This 
command will only function if you have already issued a Calibrate Begin (CB) 
command.  Note that the cell should not have any load or torque on it for this command 
to operate correctly.  The format is CS(shunt value in Lb or LbI)#.  The example shows 
calibration of a load cell using a shunt value of 553.26 Lb. 
 
Example: 
@123CS553.26# 
Acknowledge: 
@123 Calibrate Shunt Command - Reading... 
          [10 seconds delay] 
@123 Calibrate Command - Reading for Shunt Check... 
          [10 seconds delay] 
@123 Calibrate Command Completed 
     Ch A = S/N 123456, 1000.0 Lb      , 4.50014 mV/v,  
            10.00 V   , Cal on Oct27-99, 553.26 Lb  Shunt 
     unused S/N  63220, 100.00 Lb      , 2.99984 mV/v,  
            10.00 V   , Cal on Oct27-99, 49.381 Lb  Shunt 
     Ch B = S/N  89991, 1000.0 Lb      , 4.50015 mV/v,   
             5.00 V   , Cal on Oct27-99, 486.45 Lb  Shunt 
 
   Calibrate Counts-Per-Inch Command (CI) 
This command has two subcommands, a view command (CIV) and a set command 
(CIS). 
M O D E L   9 8 4 0                                                                                                                                                          P G   4 2  
P U B .
 
2 8 5 6 - 1 6