Epson 9840 User Manual

Page of 59
I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
@123LE 
Acknowledge: 
@123 Limit Setup Command Canceled 
Limit Reset Command (LiR) 
The Limit Reset (LiR) command is used to manually reset a latching limit that has been 
activated.  The format is L(limit number)R. 
Example: 
@123L1R 
Acknowledge: 
@123 Reset Limit 1 
System Options Command (O) 
The system Options command has 8 subcommands that are used to view or set the 
options that appear on the system options menu. Option View (OV) displays all the 
settings in the system options menu, Option Printer (OP) is used to set the printer 
baudrate, Option Auto ID (OI) is used to turn auto-identification of load cells on or off, 
Option Auto Zero (OZ) is used to turn auto-zeroing of load on or off, Option Com Address 
(OA) is used to change the communications address of the unit, Option Baudrate (OB) is 
used to change the communications baudrate, Option Linefeed (OL) is used to turn the 
auto line feed option on or off, and OE which is used to turn the end of transmission 
character on or off.  Each of these options is described below. 
    Option View Command (OV) 
The Option View (OV) command returns a list of the current settings of all the system 
options.  The format is OV. 
Example: 
@123OV 
Acknowledge: 
@123 Printer Baud Rate is 9600 
Auto Identify is off 
Auto Zero Channel A is off 
Auto Zero Channel B is off 
Com Address is 123 
Com Baud Rate is 9600 
Com Line Feed is off 
Retain Tare is off 
Freeze Display is off 
RS232 End of Transmission Character is off 
    Option Printer Baudrate Command (OP) 
The Option Printer baudrate (OP) command is used to select the printer baudrate.  The 
format is OP(code) where the code is from the table shown at the right. 
M O D E L   9 8 4 0                                                                                                                                                          P G   4 5  
P U B .
 
2 8 5 6 - 1 6