Series 
 
 
DAC Ranges 
The output range of the DaqBoard/500 DACs [DAC0 and DAC1] can be independently set to either 
±10 V, or 0 to 10 V. 
The following table indicates the maximum resolution for each range. Note that resolution is not accuracy. 
Resolution defines the minimum definable voltage increment. Absolute DC accuracy and relative accuracy 
define exactly how close the actual voltage output will be to the expected output. 
 
DAC OUTPUT 
DaqBoard/500 DAC Range Resolution 
Range 
Configuration 
Full Scale 
Range 
Microvolt 
Resolution 
BIPOLAR 
± 10.00 V 
305.600 
µV/bit 
UNIPOLAR 
0 to 10.00 V 
152.800 
µV/bit