Controlling Analog, Digital, and Timer Outputs 
Each setpoint can be programmed with an 8-bit digital output byte and corresponding 8-bit mask byte. 
When the setpoint criteria has been met, the P2C digital output port can be updated with the given byte and 
mask.  Alternately, each setpoint can be programmed with a 16-bit DAC update value, any one of the 4 
DAC outputs can be updated in real time.  Any setpoint can also be programmed with a timer update value. 
In hysteresis mode each setpoint has two forced update values.  Each update value can drive one DAC, one 
timer, or the P2C digital output port.  In hysteresis mode the outputs do not change when the input values 
are inside the window
.  There is one update value that gets applied when the input values are less than the 
window and a different update value that gets applied when the input values are greater than the window. 
Update on True and False uses two update values.  There is one update value that gets applied when the 
specified criteria is met (True) and a different update value that gets applied when the specified criteria is 
not met (False).  The update values can drive DACs, P2C, or timer outputs. 
 
Example:  Setpoint Detection on a Totalizing Counter 
In the following figure Channel 1 is a counter in totalize mode.  Two setpoints are used to define a point of 
change for Detect 1 as the counter counts upward.  The detect output will be high when inside the window 
(greater than Limit B (the low limit) but less than Limit A (the high limit).  In this case, the Channel 1 
setpoint is defined for the 16 lower bits of channel 1’s 32-bit value.   The P2C digital output port could be 
updated on a True condition (the rising edge of the Detection signal).  Alternately, one of the DAC output 
channels, or timer outputs, could be updated with a value. 
 
 
 
 
 
 
Limit A 
Limit B 
 
 
 
Detection
 
 
Channel 1 in Totalizing Counter Mode, Inside the Window Setpoint 
 
The detection circuit works on data that is put into the acquisition stream at the scan rate.   This data is 
acquired according to the pre-acquisition setup (scan group, scan period, etc.) and returned to the PC.  
Counters are latched into the acquisition stream at the beginning of every scan.  The actual counters may be 
counting much faster than the scan rate and therefore only every 10
th
, 100
th
, or n
th
 count will show up in the 
acquisition data.  Therefore it is possible to set a small detection window on a totalizing counter channel 
and have the detection setpoint “stepped over” since the scan period was too long. Even though the counter 
value stepped into and out of the detection window, the actual values going back to the PC may not.  This 
is true no matter what mode the counter channel is in.   
6-4   Setpoint Configuration for Output Control 