Adding Resistor/Capacitor Filter Networks 
WARNING 
Disconnect the DBK215 from power and signal sources prior to installing capacitors or 
resistors. 
 
CAUTION 
 
 
 
Ensure wire strands do not short power supply connections to any terminal potential.  
Failure to do so could result in damage to equipment. 
Do not exceed maximum allowable inputs (as listed in product specifications).  There 
should never be more than 30 V with reference to analog ground (AGND) or earth 
ground. 
You must provide strain-relief (lead slack) to all leads leaving the module.  Use tie-wraps 
[not included] to secure strain-relief. 
Always connect the CHASSIS terminal to earth ground.  This will maximize static 
protection.
 
If a channel is not associated with a DBK expansion option you can install a customized RC filter network 
to improve the signal-to noise ratio, assuming that an unacceptable level of noise exists.  DBK215’s 
internal board includes silk-screened sockets for installing RC filter networks.  The following table 
contains values that are typical for RC filter network components. 
Typical One-Pole Low Pass Filter 
Values 
for DBK215 
Ohms µF Hertz 
(-3dB) 
kHz 
(-3dB) 
510 1 
312 
0.31 
510 0.47 664 
0.66 
510 0.22 
1419 
1.42 
510 0.1 
3122 
3.12 
510 0.047 6643 6.64 
510 0.022 
14192 
14.19 
510 0.01 
31223 
31.22 
510 0.0047 66431 66.43 
470 0.0033 
102666 
102.67 
 
Do not use RC filters in conjunction with additional DBK expansion 
accessories. 
 
An Example of Customer-Installed 
 Capacitors and Filters for RC Networks 
In this example Channels 0 and 8 are shown as Single-Ended.   
Channel 1 is Differential, i.e., using 1H and 1L (channel High and Low). 
 
The following three notes pertain to the above figure
Note 1:  The 3 horizontal capacitors [as oriented in the illustration] are optional filter capacitors. 
Note 2:  The vertical capacitor [as oriented in the illustration] is an optional isolation capacitor used for the 
reduction of Differential noise.  Such capacitor placement is not used in Single-Ended applications. 
Note 3:  If installing filter resistors, carefully drill out the indicated centers with a 1/16 inch drill-bit.  Otherwise 
the resistor will be short-circuited. 
 
Prior to installing RC components, review the previous Warning and Caution 
statements, then read over the following information regarding resistors and 
capacitors. 
 