Fluke A70M User Manual

Page of 148
 
3. OPERATION 
 
Page 51 
© 2007 DH Instruments, a Fluke Company 
The over pressure function causes the measured pressure display to flash.  The 
overpressure function also logs the time and date of the overpressure condition in 
both user and factory logs to assist in incident diagnosis (see Section 3.5.7.5). 
To recover from an overpressure condition, cycle RPM4 power.  Before cycling 
power, be sure to correct the condition that caused the overpressure. 
3.4.5 
<5AT
EST
 PURPOSE 
To view data collected while running an AutoTest and/or to view and edit File AutoTest 
definition files. 
See Section 3.3.10, PRINCIPLE for information on the AutoTest function. 
 OPERATION 
To access the ATest menu select [SETUP], <5ATest>.  The display is: 
Select  <1data> to view AutoTest results data (see 
Section 3.4.5.1). 
Select  <2file> to view or edit an AutoTest definition 
file (see Section 3.4.5.2). 
AutoTest: 
1data 2file 
 
3.4.5.1 
DATA 
 PURPOSE 
To view the data files that are created when an AutoTest is run. 
 PRINCIPLE 
RPM4 supports AutoTest automated tests (see Section 3.3.10).  When an 
AutoTest is run, the RPM4 pressure measurement accepted by the operator at 
each point is logged.  The logged pressure measurements and an identifying 
header are stored in a data file.  The data file can be viewed immediately 
following the last point of the sequence by selecting <1data>.  In addition, the 
last ten data files recorded can be viewed anytime by selecting [SETUP]
<5ATest><1data>
Data File Protocol 
Up to ten data files are buffered, ten at a time, in RPM4 memory.  When a new 
AutoTest run is completed, its data file goes to the back of the queue and the 
oldest data file is deleted. 
Data files are identified by a header whose first line is the full scale of the DUT 
and the number of test points in the AutoTest.  The second line is the date 
(YYYYMMDD) and time (HH:MM:SS) at which the last point of the sequence 
was accepted. 
Following the data file header, each test point is recorded including the pressure 
reading of the RPM4 with head, AutoZ and range status as well as an indication 
of whether the reading was out of tolerance and the test point number over the 
total number of test points in the Sequence. 
 OPERATION 
To access the AutoTest results data file viewing function press [SETUP]
<5ATest><1data>
The first display is the data file header which can be used to toggle through 
data files: