being AutoZeroed.  Given the high pressure range of the RPM4 Q-RPTs, adequate 
uncertainty is easily available from a variety of digital barometers or, for the Hi Q-RPT, by the 
Lo Q-RPT. 
In absolute measurement mode, with AutoZ ON, absolute pressure is calculated as: 
P
abs
 = P
u
 - P
offset,A
 
In absolute measurement mode, running the AutoZ function to update the AutoZ value 
(P
offset,A
), is initiated by the operator using [AutoZ] (see Section 3.3.9.1). 
 
AutoZ in gauge measurement mode, dynamic compensation for atmospheric pressure 
The RPM4 A70M/A20M-AF Q-RPTs are intrinsically absolute but they are also used in gauge 
measurement mode (difference from atmosphere).  Gauge measurement mode is achieved 
by subtracting the value of atmospheric pressure from the Q-RPT’s absolute reading using 
AutoZ and by dynamically compensating for changes in atmospheric pressure between 
opportunities for AutoZ to execute (see Section 3.2.2, 
 PRINCIPLE). 
In gauge measurement mode, the value of P
std,0
 is always zero gauge pressure.  Zero gauge 
pressure, by definition, is applied to the Q-RPT when it is vented to atmosphere.  Gauge 
pressure is the measured absolute pressure, P
u
, minus P
offset,G
P
gauge
 = P
u
 - P
offset,G
 
When AutoZ is ON, dynamic compensation for atmospheric pressure is also applied to 
compensate for changes in atmospheric pressure as measured by RPM4’s on-board barometer 
(
ΔP
atm
) between AutoZ updates (see Section 3.2.2).  The measured gauge pressure is 
calculated using 
ΔP
atm
 to correct the value of P
offset.G
P
gauge
 = P
u
 - P
offset.G
 - 
ΔP
atm
 
In gauge measurement mode, running AutoZ is initiated by the operator pressing [AutoZ] 
(see Section 3.3.9.1) when the RPM4 is vented.  This assures regular updating of the value 
of P
offset,G
 and P
atm,0
.  
ΔP
atm
 is continuously updated automatically. 
 
  When gauge mode is first activated, the value of P
offset.G
 subtracted from the absolute 
pressure reading to achieve gauge pressure is standard atmosphere of 101.325 kPa.  This can 
cause a large zero error when gauge mode is activated if atmospheric pressure is significantly 
different from standard atmospheric pressure.  When gauge mode is activated, always vent the 
RPM4 Q-RPT and press [AutoZ] to run AutoZ and obtain an actual value of 
P
offset.G
 . 
 
AutoZ ON/OFF 
The AutoZ function can be turned ON and OFF separately for gauge and absolute 
measurement modes.  Table 5 summarizes the effect of AutoZ ON and OFF.  Generally, 
AutoZ should always be left ON. 
Table 5.  AutoZ ON and OFF 
MEASUREMENT 
MODE 
AutoZ 
STATUS 
P
offset
 
APPLIED 
ΔP
atm
 
APPLIED 
AutoZ 
ROUTINE 
RUNS 
ON YES 
When initiated 
by operator 
Absolute 
OFF NO 
Not Applicable 
Not available 
ON YES 
Gauge  
OFF 
YES 
NO 
When initiated 
by operator