The INTERNAL menu choices include: 
<1baro> 
View the real time output of the RPM4 on-board barometer (see Section 
3.5.7.1).  
<2ReadRt> 
Turn ON and OFF RPM4’s automated, rate of change dependent, reading 
and integration and display rate feature (see Section 3.5.7.2). 
<3RPT2x> 
Feature not used in this RPM4 model. 
<4lo vnt> 
Feature not used in this RPM4 model. 
<5log> 
View the RPM4 incident log (see Section 3.5.7.5). 
3.5.7.1 
<1BARO> 
 PURPOSE 
To view the value of atmospheric pressure as measured by the RPM4 on-board 
barometer. 
 PRINCIPLE 
RPM4 A70M/A20M-AF is equipped with an on-board barometer.  The atmospheric 
pressure measurements made by the on-board barometer are used for dynamic 
compensation of atmospheric pressure when making gauge pressure measurements 
(see Sections 3.2.2, 3.3.3, 
 PRINCIPLE). 
 
  See Section 5.4,4 Figure 13 and Section 5.4.5, Figure 14 for a photo and 
schematics showing the position of the on-board barometer in RPM4. 
 
  The on-board barometer is a low accuracy sensor used only for measuring 
small changes in atmospheric pressure over short periods of time 
 
(see Section 3.2.2).  RPM4 measurement uncertainty does not depend on the 
measurement uncertainty of the on-board barometer. 
 
 OPERATION 
To view the current reading of the on-board barometer press [SPECIAL]
<7internal>,  <1baro>.  The display is in the active pressure unit of measure  
(see Section 3.3.2).  The display resolution is fixed to 0.1 Pa, or its equivalent. 
3.5.7.2 
<2R
EAD
R
T
 PURPOSE 
To turn ON and OFF RPM4’s automated, rate of change dependent, reading 
integration and display rate feature. 
 PRINCIPLE 
To obtain maximum resolution from RPM4 Q-RPT pressure measurements, an 
integration time of about 1.2 second per reading is used.  In most situations, 
maximum precision is needed when pressures are stable so a relatively slow 
display update rate presents no disadvantage.  However, when pressure is 
changing quickly, more rapid pressure updates are usually more important than 
obtaining maximum precision on individual readings.  The RPM4 read rate 
function automatically adjusts pressure measurement integration time depending 
on the rate of change of pressure.  When pressure is changing rapidly, reading 
rate is increased.  When pressure is evolving slowly, reading rate is decreased 
and maximum precision is obtained.