Agilent Technologies DC140 User Manual

Page of 66
 
 
User Manual: Family of 8-bit Digitizers 
 
Page 31 of 66 
Model  
   Agilent # 
Bandwidth 
into 50 
Ω 
Bandwidth/ 
Capacitance 
into 1 M
Ω 
Minimum 
Rise Time 
BW Limiter 
selections 
Maximum 
Offset 
Other 
particularities 
DC241A 
  U1064A 
950 MHz  
( 1 GHz 
typical) 
300 MHz 
typical 
14 pF 
0.35 ns 
(50
Ω) 
(1.2 ns 
typical      
1 M
Ω) 
20, 200, 700 MHz 
(see remark below) 
5 V (50
Ω) 
20 V
 
(> 0.5 V 
FS & 1M
Ω) 
200 V
 
(> 5V 
FS & 1M
Ω)
   
DC271 FAMILY 
Max FS = 50 V    
(1 M
Ω) 
DC265 
  U1063A 
150 MHz 
150 MHz 
11 pF 
2.3 ns 
NA 
20 V 
BW
50
 = 90 MHz 
DC270 
  U1063A 
250 MHz 
250 MHz 
11 pF 
1.4 ns 
NA 
20 V 
BW
50
 = 90 MHz 
DC271 
  U1064A 
1 GHz 
NA 
0.35 ns 
20, 200, 700 MHz 
5 V 
DC271 FAMILY 
DC271A 
  U1064A 
DC271AR 
950 MHz  
( 1 GHz 
typical) 
300 MHz 
typical 
14 pF 
0.35 ns 
(50
Ω) 
(1.2 ns 
typical      
1 M
Ω) 
20, 200, 700 MHz 
(see remark below) 
5 V (50
Ω) 
20 V
 
(> 0.5 V 
FS & 1M
Ω) 
200 V
 
(> 5V 
FS & 1M
Ω)
   
DC271 FAMILY 
Max FS = 50 V    
(1 M
Ω) 
The AR has only 2 
channels 
For 1 M
Ω coupling in the DC2x1A/AR the 700 MHz Bandwidth Limiter cannot be used. Furthermore, for FS gain > 
5V the 200 MHz Bandwidth Limiter is always active. 
3.2.6.  Input Voltage and Offset 
The input channel provides a fully programmable amplifier with variable input voltage and offset. Full Scale (FS) 
input voltages are selectable from 50 mV to 5 V (except as shown above) in a 1, 2, 5 sequence. Care should be taken 
to select an input voltage range that will allow the signal to be recorded using as much dynamic range of the digitizer 
as possible. The Variable Offset is programmable in the range of  ±2 V when using an FS Input Voltage setting of 
500 mV or below, increasing to ± Maximum Offset for FS settings above 500 mV. The raw 8-bit ADC data values 
are in the range [-128,+127] with the first and last values reserved for underflow and overflow respectively. The 
midpoint value, 0, of the range corresponds to the negative of the offset voltage. Thus the Full Scale Range (FSR) 
goes from  
–Offset Voltage  – (FS/2)  
 
to  
–Offset Voltage  + (FS/2) 
Signals going outside of the FSR will be clipped and data values for the clipped portion of a signal should be 
regarded as erroneous. 
The maximum input voltage for 50 
Ω  input impedance is ±5 V. The maximum input for 1 MΩ  input impedance is 
±100 V (dc + ac) except for the DC2x1A/AR models where it is ±300 V (dc + ac). 
3.2.7.  Vertical Resolution 
The digitizers described in this manual use an ADC system with 8 bits of vertical resolution (256 levels). The 
dynamic range of the ADC covers the Full Scale Range (FSR) of the Input Voltage setting. For example, if the Input 
Voltage is set to 1 V then the ADC resolution is equivalent to 3.91 mV. To obtain the best dynamic range from the 
ADC care should be taken to ensure that the input signal varies over more than 50% of the Input Voltage FSR 
setting. 
3.2.8.  DC Accuracy and Linearity 
The DP and DC Series digitizers use low noise front-end electronics in order to ensure voltage measurement is made 
with accuracy and precision. DC voltage accuracy, at 0 V offset, is better than ±2% (±1% typical) of the input 
voltage full scale. The differential linearity is better than ±0.7 LSB ( ±0.8 LSB for DC135/DC140 digitizers  and 
±0.9 LSB for other DC271-FAMILY digitizers).