Appendix B: Algorithms
TDS 500C, TDS 600B, & TDS 700C User Manual
B–3
LowRef — the waveform low reference level. Used in fall and rise time
calculations. Typically set to 10%. You can set it from 0% to 100% or to a
voltage level.
calculations. Typically set to 10%. You can set it from 0% to 100% or to a
voltage level.
Mid2Ref — the middle reference level for a second waveform (or the second
middle reference of the same waveform). Used in delay time calculations.
Typically set to 50%. You can set it from 0% to 100% or to a voltage level.
middle reference of the same waveform). Used in delay time calculations.
Typically set to 50%. You can set it from 0% to 100% or to a voltage level.
The oscilloscope also measures several values itself that it uses to help calculate
measurements.
measurements.
RecordLength — is the number of data points in the time base. You set it with
the Horizontal menu Record Length item.
the Horizontal menu Record Length item.
Start — is the location of the start of the measurement zone (X-value). It is 0.0
samples unless you are making a gated measurement. When you use gated
measurements, it is the location of the left vertical cursor.
samples unless you are making a gated measurement. When you use gated
measurements, it is the location of the left vertical cursor.
End — is the location of the end of the measurement zone (X-value). It is
(
(
– 1.0) samples unless you are making a gated measurement.
When you use gated measurements, it is the location of the right vertical cursor.
Hysteresis — The hysteresis band is 10% of the waveform amplitude. It is used
in
in
, , and calculations.
For example, once a crossing has been measured in a negative direction, the
waveform data must fall below 10% of the amplitude from the
waveform data must fall below 10% of the amplitude from the
point
before the measurement system is armed and ready for a positive crossing.
Similarly, after a positive
Similarly, after a positive
crossing, waveform data must go above 10%
of the amplitude before a negative crossing can be measured. Hysteresis is useful
when you are measuring noisy signals, because it allows the oscilloscope to
ignore minor fluctuations in the signal.
when you are measuring noisy signals, because it allows the oscilloscope to
ignore minor fluctuations in the signal.
Other Variables