ANCB Description, 
Continued 
 
Jumper J5 Settings 
The ANCB jumper settings for J5 are factory set and dependent upon the 
product the ANCB will be used in. 
Product Type 
ANCB J5 Jumper Settings 
Advance Network Gateway 
Ethernet (ANG-EN) 
Pins 3&4 are shorted, switch in 25-
ohm position. 10Base2 connector is 
not normally used (not connected) 
Advance Network Gateway 
DataNET (ANG-DN) 
Pins 1&2 are shorted, switch in 50-
ohm position. 10Base2 connector is 
used as an end point for crossover 
cable. 
Advance Plus Analyzer 
Pins 1& 2 are shorted, switch in 50-
ohm position. 10Base2 connector is 
used as an end point for crossover 
cable. 
Maxum Analyzers with ANCB/DN 
Installed 
Pins 3&4 are shorted, switch in 25-
ohm position. 10Base2 connector is 
not normally used (not connected) 
 
More Information 
See ANCB LEDs Description and Operation, page 49
48  
  Troubleshooting 
2000592-001