Gossen R2600 User Manual

Page of 40
GOSSEN METRAWATT GMBH
7
Following, the master may only become active again, when
– it receives a reply telegram from the addressed equipment and the specified wait time after a reply telegram (t aw) has elapsed, or
– the maximum specified response delay time (t
av
) has elapsed.
Within a telegram, pauses of limited duration (t
zv
 = character delay time) may occur between 2 character transmissions.
1.5 Time action
Ready to send/receive after turn-on
t
ber
approx. 5 s
Character delay time (R2600 transmitter)
t
zvs
<
3 ms
Character delay time (master)
t
zvm
<
500 ms
Response delay time (R2600 transmitter)
10 ms <
t
av
< 100 ms
Wait time after R2600 response (master)
t
aw
>
10 ms 
Time
Master
sends
Equipment
responds
t
ber
approx. 5 s
t
av
10 ...
100 ms
t
zvs
< 3 ms
Master
sends
t
aw
> 10 ms
t
zvm
< 500 ms