Emerson Network Router User Manual

Page of 110
10
ProLink
®
 II Software for Micro Motion
®
 Transmitters
Installation and Setup
Table 2-2
Communication protocols and wiring methods for IFT97xx and RFT97xx transmitters
Transmitter type
Wiring method
IFT9701 
IFT9703
RFT9712
RFT9739
Wiring 
method #
HART protocol
Bell 202 physical layer
• Temporary or hard-wired connection to transmitter or 
multidrop network
1
• Temporary connection to field-mount transmitters
2
• Temporary connection to rack-mount transmitters
3
RS-485 physical layer
• Temporary or hard-wired connection to transmitter or 
multidrop network
5
Modbus protocol 
(RS-485 physical layer)
• Temporary or hard-wired connection to transmitter or 
multidrop network
6
Table 2-3
Communication protocols and wiring methods for Model 1500/2500, Model 1700/2700, and 
Series 3000 transmitters
Transmitter type
Wiring method
Model 
1500/2500 
Model 
1700/2700
Series 
3000
Wiring 
method #
HART protocol
RS-485 physical layer
• Temporary or hard-wired connection to RS-485 
terminals
AN
5
Bell 202 physical layer
• Temporary or hard-wired connection to primary mA 
output or multidrop network
(1)
(1) Except Model 1500 transmitter with the Filling and Dosing application. The Model 1500 transmitter with the Filling and Dosing 
application does not support HART communication.
AN
CIO
1
• Temporary or hard-wired connection to primary mA 
output or multidrop network
IS
4
Modbus protocol 
(RS-485 physical layer)
• Temporary or hard-wired connection to RS-485 
terminals
AN
6
• Temporary connection to service port
7