7-10
 
Operating Concepts
Output Power
NOTE
After measurement calibration, you can change the power within a range and 
still maintain nearly full accuracy. In some cases better accuracy can be 
achieved by changing the power within a range. It can be useful to set 
different power levels for calibration and measurement to minimize the 
effects of sampler compression or noise floor.
If you decide to switch power ranges, the calibration accuracy is degraded and 
accuracy is no longer specified. However, the analyzer leaves the correction 
on.
The annotation 
C
 will be displayed whenever you change the power after 
calibration.
Power Coupling Options 
There are two methods you can use to couple and uncouple power levels with the analyzer:
• channel coupling
• port coupling
By uncoupling the channel powers, you effectively have two separate sources. Uncoupling 
the test ports allows you to have different power levels on each port.
Channel coupling 
 toggles between coupled and uncoupled channel power. With the 
channel power coupled, the power levels are the same on each channel. With the channel 
power uncoupled, you can set different power levels for each channel. For the channel 
power to be uncoupled, the other channel stimulus functions must also be uncoupled 
(
).
Test port coupling 
 toggles between coupled and uncoupled test ports. With the test 
ports coupled, the power level is the same at each port. With the ports uncoupled, you can 
set a different power level at each port. This can be useful, for example, if you want to 
simultaneously perform a gain and reverse isolation measurement on a high-gain 
amplifier using the dual channel mode to display the results. In this case, you would want 
the power in the forward direction (S
21
) much lower than the power in the reverse 
direction (S
12
).
CH PWR [COUPLED] 
COUPLED CH OFF 
PORT PWR [COUPLED]