3. PRECAUTIONS 
To get the optimum performance from the balance and acquire accurate weighing data, note the 
following: 
3-1 Before Use 
  Install the balance in an environment where the temperature and humidity are not excessive. 
The best operating temperature is about 20°C / 68°F at about 50% relative humidity. 
  Install the balance where it is not exposed to direct sunlight and it is not affected by heaters 
or air conditioners. 
  Install the balance where it is free of dust. 
  Install the balance away from equipment which produces magnetic fields. 
  Install the balance in a stable place avoiding vibration and shock. Corners of rooms on the 
first floor are best, as they are less prone to vibration. 
   The AD-4212A/B series balance responds even to very subtle 
air flow. To avoid the influence of ambient air flow or air flow 
caused by balance operation, make sure to use the dust 
guard and the breeze ring (AD-4212B-101/102 only). 
 Level the weighing unit by adjusting the leveling feet 
and confirm it using the bubble spirit level. 
  If the leveling adjustment is difficult to perform due to 
the installation conditions, perform calibration using a 
calibration weight or controlled actual sample before 
weighing. 
 Ensure a stable power source when using the AC 
adapter 
  Warm up the balance for the appropriate duration. Plug 
in the AC adapter as usual. 
• AD-4212A series balance: 30 minutes or more 
• AD-4212B series balance: one hour or more 
  Calibrate the balance before use or after having moved 
it to another location.   
  In addition, calibrate it periodically to maintain the accuracy.     
Caution 
Do not install the balance where flammable or corrosive gas is present.