 Magnetic material 
The balance uses a strong magnet as part of the balance assembly, so use much care when 
weighing magnetic materials. 
  Place a non-magnetic object such as aluminum or brass between the sample and the 
balance, also keep an appropriate distance between them while weighing. 
 
3-3 During Use 
  To minimize the affect by electrical noises, earth ground 
the weighing unit and the display unit. 
  Make each weighing gently and quickly to avoid errors 
due to changes in the environmental conditions.
 
  Do not drop things upon the weighing pan, or place a 
sample on the pan that is beyond the balance weighing 
capacity. Place a sample in the center of the weighing 
pan.
 
  Do not use a sharp instrument such as a pencil to press 
the keys. Use your finger only. 
 
 Press the RE-ZERO key before each weighing to 
prevent possible errors. In addition, a RE-ZERO signal 
can be sent using external contact input/RS-232C 
command. 
  Take into consideration the affect of air buoyancy on a sample when more accuracy is required.
 
  Keep the balance interior free of dust and foreign materials. 
 
3-4 After Use 
  Avoid mechanical shock to the balance. 
  Calibrate the balance, using a calibration weight, periodically. 
  Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or 
repair. 
  Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is 
moistened with warm water and a mild detergent. 
  Avoid dust and water so that the balance weighs correctly. Protect the internal parts from liquid 
spills and excessive dust. 
 
3-5 Power Supply 
  When the AC adapter is connected, the balance is in the standby mode if the standby indicator 
is on (refer to “4 DISPLAY SYMBOLS AND KEY OPERATION”). This is a normal state and 
does not harm the balance. For accurate weighing, plug in the AC adapter and warm up the 
balance for the appropriate duration before use.