A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
31
 
 
 
 
Class Item 
and 
Parameter 
Description 
 0 
Plus only 
Displayed value>Reference 
 1 
Minus only 
Displayed value<Reference 
ap-p
 
Auto print polarity
 
 2 
Both 
Regardless of displayed value 
 0 
10 digits 
 1 
100 digits 
ap-b
 
Auto print difference
 
 2 
1000 digits 
Difference between reference 
value and displayed value 
 0 
Not used 
 1 
Stores unit mass in counting mode 
 2 
Stores weighing data 
 3 
Stores calibration data 
data
 
Data memory
 
 4 
Stores upper/lower limit values
 
Related items:   
prt, int, d-no, 5-td, info 
 0 
Every measurement 
 1 
2 seconds 
 2 
5 seconds 
 3 
10 seconds 
 4 
30 seconds 
 5 
1 minute 
 6 
2 minutes 
 7 
5 minutes 
int
 
Interval time
 
 8 
10 minutes 
Interval time in the interval 
memory mode when using prt 3, 
data 2 
 0 
No output 
d-no
 
Data number output
 
 1 
Output 
Refer to "14. DATA MEMORY". 
 0 
No output 
 1 
Time only 
 2 
Date only 
5-td 
 
Time/Date output
 
 3 
Time and date 
Selects whether or not the time 
or date is added to the weighting 
data. For details, refer to "10-9 
Clock and Calendar Function". 
 0 
No output 
5-id
 
ID number output
 
 1 
Output 
Selects whether or not the ID 
number is output. 
 0 
No pause 
pU5e
 
Data output pause
 
 1 
Pause (1.6 seconds) 
Selects the data output interval. 
 0 
Not used 
at-f
 
Auto feed
 
 1 
Used 
Selects whether or not automatic 
feed is performed. 
 0 
No output 
  1
 
AD-8121 format 
info 
 
GLP output
 
  2
 
General data format 
Selects GLP output method. 
For how to set time and date to 
be added, refer to “10-9 Clock 
and Calendar Function”. 
 0 
Not used 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
dout 
Data output
 
 
Not displayed 
when BCD 
output (OP-01) 
is installed. 
 
 
ar-d
 
Zero after output
 
 1 
Used 
Adjusts zero automatically after 
data is output. 
 
 
Factory setting 
Note: “Digit” is a unit of minimum weighing value