A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
38
L A B - 0 1 2 3
CR LF
2 0 0 4 / 0 7 / 0 1
CR LF
CR LF
1 2 : 3 4 : 5 6
LAB-0123, No,012, 2004/07/01, 12:34:56, ST,+000.1278,~~g<CR><LF>
CR LF
S T , + 0 0 0 . 1 2 7 8 ,
  g
CR LF
O L , + 9 9 9 9 9 9 9 E + 1 9 ,
g
 
CSV format 
5if type 5 
  This format separates the data of A&D standard format and the unit by a comma ( , ). 
  This format outputs the unit even when the data is overloaded. 
  When a comma ( , ) is selected for decimal point, separators are set to semicolon ( ; ). 
  When the ID number, data number, time and date are added at "Data output (dout)" of the 
function table, outputs ID number, data number, date, time and weighing data in this order and 
separates each item by a comma and treats all the items as one group of data.
 
 
 
 
 
 
 
 
10-7  Description of the Data Format Added to the Weighing Data 
 
ID number 
dout 5-id 1 
The number to identify a specific balance.   
 
This format consists of eight characters excluding the terminator.
   
 
 
Data number    dout d-no 1 
This format outputs the data number just before the data is transmitted using the RS-232C interface.   
  This format consists of six characters excluding the terminator. 
  When CSV format (5if type 5) is selected, the period ( . ) is replaced with a comma ( , ). 
 
 
 
 
Date  dout 5-td 2 or 3 
  The date output order can be changed in " Time/Date output (5-td)" and "Clock (Cl adj)".  
  The year is output in a four-digit format.   
 
 
Time  dout 5-td 1 or 3 
  This format outputs time in 24-hour format. 
 
 
 
 
Data number 
Terminator 
ID number 
Data number 
Date 
Time 
Weighing data