A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
41
 
Confirming the date 
6  The current date is displayed with all the digits blinking. 
  To change the display order of year (y), month (n) 
and day (d), press the MODE key. The date is 
output in the order as specified. 
  
When the date is correct and the operation is to be 
finished, press the CAL key and proceed to step 8. 
  
When the time is to be confirmed again, press the 
SAMPLE key and go back to step 4. 
  
When the date is not correct and is to be changed, 
press the RE-ZERO key and proceed to step 7. 
Note 
The year is expressed using a two-digit format. For 
example, the year 2001 is expressed as “01”. 
 
Setting the date (with part of the digits blinking) 
7  Set the date using the following keys. 
SAMPLE key 
To select the digits to change the 
value. The selected digits blink. 
RE-ZERO key 
To increase the value by one. 
MODE key 
To decrease the value by one. 
PRINT key 
To store the new setting, display 
1
end
1
 and go to step 8. 
CAL key 
To cancel the new setting and go 
to step 8. 
Quitting the operation 
8  The balance displays the next menu of the function 
table. Press the CAL key to exit the clock and calendar 
function and return to the weighing mode. 
Notes 
 
Do not enter invalid values such as a non-existing date when setting the time and date. 
 
When the clock backup battery has been depleted, the balance displays    rtc  pf  . Under 
this condition, press any key and set the time and date. The dead battery only affects the 
clock and calendar function. Even so, the function works normally as long as the AC 
adapter is connected to the balance.