15. PROGRAMMABLE-UNIT 
(AD-4212A only)
 
This is a programmable unit conversion function. It multiplies the weighing data in grams by an 
arbitrary coefficient set in the function table and displays the result.   
The coefficient must be within the range between the minimum and maximum shown below. If the 
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient 
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory. 
Model 
Minimum coefficient 
Maximum coefficient 
AD-4212A-100 10000 
AD-4212A-200/600/1000 
0.000001 
1000 
 
 
Operation 
1  Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key. 
2  Press the SAMPLE key several times to display    mlt
Cadj
 . 
3  Press the PRINT key. The balance enters the mode to confirm or set the coefficient. 
Confirming the coefficient 
4  The current coefficient is displayed with the first digit blinking.   
  
When it is not to be changed, press the CAL key and proceed to step 6. 
  
When it is to be changed, press the RE-ZERO key 
and proceed to step 5. 
Setting the coefficient 
5  Set the coefficient using the following keys. 
SAMPLE key 
To select a digit to change the 
value. The selected digit blinks. 
RE-ZERO key 
To change the value. 
MODE 
key 
To change the decimal point 
position. 
 
Each time the switch is pressed, 
the decimal point position 
changes as follows: 
 
 
 
PRINT key 
To store the new setting, display 
1
end
1
 and go to step 6. 
CAL key 
To cancel the new setting and go 
to step 6. 
Selecting the programmable-unit (        ) 
6 The balance displays  Unit . Press the PRINT key to enter the unit selection mode. Select the 
programmable-unit (        ) as described in “5-2 Changing the Unit”.   
7  Press the CAL key to exit the programmable-unit function and return to the weighing mode. 
Using the function 
Perform weighing as described in “6-1 Basic Operation (Gram Mode)”. After weighing, the balance 
displays the result (weighing data in grams x coefficient).