A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
71
19.  BCD OUTPUT (OP-01) 
Weighing data will be output in BCD, in sync with the display refreshing. In addition, the polarity (+/-) 
and balance condition (stable/not stable, overload(positive/negative)) will be output. 
The strobe signal allows inputting data easily. Inputting BUSY will hold data or prevent data that is 
being output from being rewritten. The logic of weighing data, status, and strobe signal can be 
switched individually in the function table. 
Contact input is available for RE-ZERO and ON/OFF. (The same operations as pressing the 
RE-ZERO and ON/OFF keys.) 
 
19-1  Connector Pin Nos. and Specifications 
I/O connector located on the rear of the display unit 
 
 
 
 
 
 
Plug (provided) 
Part name 
Product number 
Manufacturer 
Over mold cover 
DX30M-50-CV 
Plug unit (soldered type) 
DX40M-50P 
Hirose Electric 
 
The products above are subject to be replaced with the equivalent. 
 
Cable 
Wire size 
AWG #28 
Core configuration 
7/0.127 
O.D. of insulator 
0.58 
 
Use a shielded cable and connect the shield to the connector housing.